>>> 'tylan14' problem from bob -- Tue Jun 27 15:53:24 1989 <<<

seh reports spiking oon the aluminum he annealed in tylan14. It 
appears that the temp calibration may be off. Tom Booth will run
calibration and then we will check this tube.

>>> 'tylan14' problem from tom -- Thu Jun 29 11:44:05 1989 <<<

Problem:   User reports his Al was spiked, possibly due to temp being
           out of calibration.
Cause:     Unknown. Will check tube.

Solution:  Checked calibration. TC's read correctly, 400+-2 C, according
           to TC calibration device. A calibrated signal sent into the
           TC port gives the right reading on the tycom. The user says
           the tube over shot to 440 C during his run. This MIGHT have 
           caused his spiking. It is not likely the tube is off at 440 C
           and good at 400 C. We have a procedure to reduce the overshoot
           to <10 C so this can be avoided in the future for interested 
           users.  Tube is up.

>>> 'tylan14' problem from evan -- Wed Dec 20 16:31:44 1989 <<<

The boat loader has been moved to tylan16.  tylan14 is down until the
broken boat loader fro tylan16 is repaired and installed.

>>> 'tylan14' problem from evan -- Mon Jan  8 16:54:31 1990 <<<

problem: boat loader module disfunctional
cause: a.c. wiring harness had an open lead
solution: the a.c. harness was caught between the module and the frame. The
lead was smashed and weakened. The lead was spliced back together and the
harness tiewrapped away from the frame.
tylan14 is up
.`

>>> 'tylan14' problem from jchen -- Sun Mar 25 22:37:36 1990 <<<


Serious problem. It shows scr error and the whole power is off.
That's why cannot even put the tube back in. Don't know the reason
for this. Was running a normal anneal program.

>>> 'tylan14' problem from evan -- Mon Mar 26 16:20:16 1990 <<<

The scr alarm is tripping because the cooling fan is not operating.
The fan will be replaced.
The boat loader is also not running in insert mode. This will be 
corrected as well.
tylan14 is down

>>> 'tylan14' problem from evan -- Tue Mar 27 11:12:17 1990 <<<

problem: scr alarm
cause: scr fan failed
solution: replaced scr fan.  also, boatloader was checked and reset.

tylan14 is up