>>> 'sink7' problem from phillip -- Thu Nov 30 08:30:24 1989 <<<

P: tank4 drain not working.
S: repaired open air line on SV13.
   Sink7 is up.
From phillip Thu Feb  1 08:59:16 1990
To: bob
Subject: sink solenoid upgrade
Cc:  $pf/sink8-p $pf/sink7-p $pf/sink6-p kellogg 


  A salesman by the name of Steve Popolizio from Elect. air tool
is coming by on 2/12/90, Monday . We will be talking/specifying/
ordering  Honeywell solenoids for this project. 
 He is also bringing an updated Elect Air Tool catalog.

(Susan, please arrange parking for him @ the East gate.
 Please apprise me when this is done. Thank you.      )


>>> 'sink7' problem from marilyn -- Fri Feb  2 07:41:55 1990 <<<

Problems: 1) rinse tank 3 does not hold water
          2) aspirator and plenum flush cycles need to be synchronized

>>> 'sink7' problem from jimmoon -- Fri Feb  2 09:26:37 1990 <<<

Phosphoric acid bath reads 150C but is actually at room
temperature.

>>> 'sink7' problem from phillip -- Fri Feb  2 09:52:57 1990 <<<

The phosphoric acid bath temp controller needs calibration.
This will be done today. 

>>> 'sink7' problem from phillip -- Fri Feb  2 11:48:09 1990 <<<

P: P acid bath does not heat according to controller readout.
S: Replaced J type tc. Calibrated controller. Sin7 is up.
   H3PO4 bath is now operational.

>>> 'sink7' problem from field -- Fri Feb 16 12:05:07 1990 <<<

heater for phosphoric acid bath doesn't seem to be working.  It 
is set to heat to 150 C and is only at 23-24 C after about 1 hour of 
operation.

>>> 'sink7' problem from phillip -- Fri Feb 16 13:26:53 1990 <<<

Replaced phosphoric acid bath thermocouple. This 
repairs the problem as reported by Leslie.

>>> 'sink7' problem from cjkim -- Thu Mar 22 00:07:25 1990 <<<

spindryer for sink7&8 keeps slowly rotate(rpm ~ 2) after normal spin.
Had to turn off the power to stop it.

>>> 'sink7' problem from cjkim -- Thu Mar 22 01:49:46 1990 <<<

The slow rotation I previously reported did not happen this time.

>>> 'sink7' problem from kpister -- Fri Mar 30 18:24:55 1990 <<<

spindryer2 stopped in mid-cycle several times.  other people were using
lots of nitrogen at the time.  wasn't able  to repeat the problem when
other nitrogen users stopped.

>>> 'sink7' fix from phillip -- Tue May  8 07:36:46 1990 <<<

Fabricated and replaced 'J' type thermocouple in
Phosphoric acid bath. The problem as reported
by Nam is fixed.