>>> 'dw' problem from hebert -- Wed Jan  3 08:34:25 1990 <<<

Problem: Dirty chamber.

Solution: I cleaned the chamber, replenished the sources (Mo, Al, SiO).
 

The Al is now in crucible #2. Crucible #3 is not being used. The Mo
source needs to be conditioned before it can be used. The cryo pump was
regenerated.

This morning the base pressure is 9.0 x 10 E-8 Torr with the copper
shroud hot.

The DW is UP.

>>> 'dw' problem from parrish -- Fri Jan 26 07:00:59 1990 <<<

system down for investigation of noises from roughing pump.

>>> 'dw' problem from parrish -- Fri Jan 26 09:38:23 1990 <<<

problem:    noises at the roughing-pump

cause:      the fan on the motor worked itself loose

solution:   removed the motor readjusted the fan, and retightened
            replaced the motor. system is up.

>>> 'dw' problem from bob -- Mon Jan 29 11:08:15 1990 <<<

The mechanical pump on the dw has problems with its motor. It has
been turned off and vented. James Parrish has been informed.

>>> 'dw' problem from parrish -- Fri Feb  2 09:25:42 1990 <<<

problem:    motor making noise

cause:      have not determined the source of the noise, other than it
            seem to be coming from the motor.

solution:   allow pump to be used temporarely, and order another motor.

>>> 'dw' problem from hebert -- Fri Feb 23 15:40:19 1990 <<<

The mechanical pump makes noise and must be turned off
after roughing the chamber.

Also, the water pump does not turn off after the e-gun and
power supply water cooling valves are turned off.

>>> 'dw' problem from bob -- Wed Mar 21 10:23:23 1990 <<<

The mechanical pump on the dw is being changed to a D16A.
Problems reported by Dave Hebert on the recirc water have
been serviced by him.
The dw will be up by 11:00am today.
.

>>> 'dw' problem from parrish -- Wed Mar 21 11:11:56 1990 <<<

problem:    pump making noises

cause:      undetermined, pending further inspection

solution:   removed pump and replaced with a spare. system is up.

>>> 'dw' problem from hebert -- Thu Apr  5 16:08:38 1990 <<<

The cryo pump dumped this afternoon. I regenerated it. I have to
regenerate this pump about once a week or every 4 pump downs.


The chamber should be He leak checked. The problem may be due
in part to a the high vac valve seal: maybe it has dirt on it
again.

>>> 'dw' fix from hebert -- Wed Apr 11 14:16:41 1990 <<<

Problem: resistance monitor bridge lead broken.

Solution: resoldered lead to bridge contact.
>>> 'dw' problem from hebert -- Thu Apr 12 07:46:51 1990 <<< [No Comment]