>>> 'sink2' problem from phillip -- Tue Jan  3 06:53:43 1989 <<<

P: Liu reports the ultrasonic cleaner is o.d.
S: Found o.k. The cleaner is working.

>>> 'sink2' problem from liang -- Thu Feb  2 22:48:52 1989 <<<


both heaters on sink2 can not be turned on.

>>> 'sink2' problem from phillip -- Fri Feb  3 08:16:25 1989 <<<

P: Both heaters fail to operate.
S: Replaced lamps in 'low liquid' alarm. For
   whatever reason these alarms went on, there
   was no way to tell they needed to be reset.
  
  Sink2 is working normally again.

>>> 'sink2' problem from liu -- Thu Mar  2 14:06:19 1989 <<<

	The "ASP" did not work.

>>> 'sink2' problem from phillip -- Thu Mar  2 14:29:36 1989 <<<

I had the water shut off to the aspirator while working in the chase.
It is back on now. The trouble is fixed.

>>> 'sink2' problem from liu -- Thu Mar  2 16:57:33 1989 <<<

	"ASP" still does not work.

>>> 'sink2' problem from phillip -- Fri Mar  3 07:40:31 1989 <<<

P: Aspirator does not work (as reported by liu).
S: Replaced a defected ball valve on the ICW supply
   to aspirator. This feature awa the sink are now working.

(Liu, I aspirated the acid you had on the sink. Your beaker is
 still there, however. pg)

>>> 'sink2' problem from magneton -- Fri Apr 28 14:22:33 1989 <<<

Aspirator is a on the slow side.  It only works if you lower the
tube's maximum height below the level of the blue plastic.

>>> 'sink2' problem from phillip -- Mon May  1 08:29:02 1989 <<<

P: Aspirator not working.
S: Replaced clogged solenoid supply air tubing tee.
   The aspirator now works properly.

>>> 'sink2' problem from bob -- Fri Aug 18 14:06:56 1989 <<<

The circuit breaker on sink2 was found in the off position for
no apparent reason. Perhaps it was bumped yesterday when the
lab wet vac was being used. It is back on.

>>> 'sink2' problem from evan -- Fri Sep 22 14:59:10 1989 <<<

Leak from vac drain fitting under sink. This is dumping our vacdrain and
so I shut off this drain.  Phill will repair this Friday morning.

>>> 'sink2' problem from phillip -- Mon Sep 25 08:30:37 1989 <<<

P: Aqua-matic drain valve DEAD.
S: Replaced drain valve defected by organic chemical
   usage at this sink. 
   This wps (and the vacuum drain system) is up.

>>> 'sink2' problem from phillip -- Thu Oct 12 14:43:44 1989 <<<

This sink is DOWN aand unusable until tomorrow

>>> 'sink2' problem from phillip -- Fri Oct 13 11:46:31 1989 <<<

Sink2 is down because of a vacuum drain problem.
It will hopefully be repaired by Monday afternoon.

>>> 'sink2' problem from phillip -- Mon Oct 16 11:49:28 1989 <<<

Replaced defective aqua-matic drain valve.
Sink2 is up and available for use.
>>> 'sink2' problem from xwu -- Sun Dec 10 18:42:58 1989 <<< [No Comment]

>>> 'sink2' problem from xwu -- Tue Dec 12 11:08:02 1989 <<<

no problem, sorry, i touched a wrong key.
xin

>>> 'sink2' problem from marilyn -- Thu Jan 25 13:26:03 1990 <<<

Problem: Ultrasonic bath appears to be dead.

>>> 'sink2' problem from phillip -- Thu Jan 25 14:01:24 1990 <<<

P: Ultrasonic bath is not working per Marilyn.
S: Plugged AC plug back into base of unit. The 
   ultrasonic unit and sink2 are up.

>>> 'sink2' problem from hsohn -- Thu Feb  1 17:44:58 1990 <<<

It seems that the monitor for DI water resistivity doesn't work or the resistivity of DI water is too low(below 0.01 megaohm).

>>> 'sink2' problem from phillip -- Fri Feb  2 10:42:56 1990 <<<

P: Resistivity meter not working per hsohn
S: This feature was not employed when the sink
   was built. The resistivity meter does not 
   work on this sink.