>>> 'sink12' problem from bob -- Mon Jan  9 14:51:04 1989 <<<

The float assembly on sink12 has been replaced. I used the last spare
float assembly to complete this repair. 
sink12 is up for use.

>>> 'sink12' problem from bob -- Thu Jan 19 14:19:33 1989 <<<

The floats on sink12 dumped the vacdrains. All floats were in their
lowest position and when the unit was moved ~1 inch the drain reset
to the proper position. Pins in the connector are a prime suspect
of failure and will be replaced.

>>> 'sink12' problem from phillip -- Fri Jan 20 06:59:06 1989 <<<

Replaced female pins in AMP connector. 
The floats failed to fail 20 0f 20 attempts.
Sink12 is working normally.

>>> 'sink12' problem from marilyn -- Tue Feb 21 08:33:56 1989 <<<

Problem: full plenum lockout

>>> 'sink12' problem from phillip -- Thu Feb 23 07:44:41 1989 <<<

P: FPL as reported by Marilyn.
S: No trouble found. The sludge that is usually present in
   this sink, and that usually causes the drain problem,is
   not a problem this time.

>>> 'sink12' problem from lyons -- Mon Apr 10 17:08:40 1989 <<<


plenum level alarm was going off, apparantly due to the 
house air pressure problems.

>>> 'sink12' problem from straub -- Sat Aug 26 10:14:25 1989 <<<

don't use this sink until further notice.  It trips all the plenum alarms.
I've notified Phil of the problem.

>>> 'sink12' problem from phillip -- Mon Aug 28 06:53:44 1989 <<<

P: Full plenum lockout. Lab drain system lost vacuum.
S: Released stuck float on sink12. Cleaned floats. 
   The emergency as reported by M. Straub on Saturday
   is fixed.

>>> 'sink12' problem from liew -- Fri Sep  1 15:38:13 1989 <<<

Full plenum lockout while developing a mask

>>> 'sink12' problem from phillip -- Tue Sep  5 06:57:28 1989 <<<


P: FPL on this sink while aptchrome running.
S: No apparent failure on this sink that caused the problem.
   The vac. drain sensor history indicates there is a problem
   with something in the vacuum make-up equipment.
   This will be explored.

>>> 'sink12' problem from marilyn -- Wed Oct  4 08:29:25 1989 <<<

Problem: Full plenum lockout; alarm was silenced, then it came
         back on and was silenced again.

>>> 'sink12' problem from phillip -- Wed Oct  4 09:23:29 1989 <<<

Inspected sink12 drain system. All appears in good working
order. No trouble found.

>>> 'sink12' problem from marilyn -- Fri Feb  2 07:27:33 1990 <<<

Problem: Possible slow plenum leak at this sink. 

>>> 'sink12' problem from phillip -- Mon Feb  5 09:47:52 1990 <<<

P: PLenum leak (?) per Marilyn.

S: Disassembled plenum to plumbing transition. Taped and 'gooped'
   all plumbing connections upon reassembly. The plenum
   no longer sporatically leaks. 

Sink12 is available for use.

>>> 'sink12' problem from marilyn -- Wed Feb 21 10:03:44 1990 <<<

Problem: full plenum lockout; alarm silenced.

>>> 'sink12' problem from bob -- Wed Feb 21 16:06:52 1990 <<<

Chlorox was added to sink12 to dissolve emmulsion residues.
This has repaired the "full plenum lockouts" on this sink in the
past. sink12 is up.

>>> 'sink12' problem from marilyn -- Fri Mar  2 13:30:12 1990 <<<

Problem: sink12 does not enable on the taurus (green light stays off.)

>>> 'sink12' problem from evan -- Fri Mar  2 16:12:33 1990 <<<

The taurus enable cable is working for sink12.  The enable light does not light
because the ac power to sink12 (and the hepa lights) is off during a repair.

>>> 'sink12' problem from phillip -- Mon Mar  5 06:50:25 1990 <<<

P: No taurus lamp lit when enabled.
S: Found power switch on sink off. Restored switch.
   Sink12 is up.