>>> 'reichert' problem from bob -- Tue Jan  3 08:24:25 1989 <<<

Problem: Motorized nose piece won't turn.
Cause:   Blown fuse
Solution: The fuse was replaced with an identical rated fuse which blew
again (50ma). A 100 ma fuse has been substituted and appears to be working.

>>> 'reichert' problem from bob -- Tue Jan  3 13:29:15 1989 <<<

This is a test of reichert problems reports.

>>> 'reichert' problem from galewski -- Mon Jan  9 21:44:51 1989 <<<

Have to watch setting the clutch too hard.  Found that the microscope
was not parcentering when going to smaller magnification.  Discovered
that the clutch was set so tight that the nosepiece would not get to
the stop when rotating clockwise.  Loosened the clutch a little and
the rotation in both directions was fine.  Mentioning this so that all
problems with the rotation are not fixed by tightening the clutch.

				Thank you,
				  Carl

>>> 'reichert' problem from bob -- Tue Jan 10 07:57:39 1989 <<<

The condenser for the transillumination light source has arrived 
and will be installed by evan. This completes the original
order. A letter has been sent to Andre Lebarsky requesting
the identification of the filter types installed, a demo of
a celan room camera, the screws to link the cubes together
and a box to house components.


>>> 'reichert' problem from richard -- Thu Jun  8 17:27:12 1989 <<<

highest magnification on reichert is not working.

the image should stay in focus when magnification is changed
but it doesn't.

>>> 'reichert' problem from evan -- Fri Jul  7 15:41:34 1989 <<<

Removed film back paper from polaroid back. Cleaned rollers.
Camera is back up.

>>> 'reichert' problem from galewski -- Tue Jul 25 17:56:55 1989 <<<

100x is out of parfocal with 50x by 30 um up.

>>> 'reichert' problem from bob -- Tue Aug  1 10:34:59 1989 <<<

Reichert has been called regarding lingering problems
with our Polyvar microscope. A week has passed since our 
last phone call without a response. Problems, such as the 
100X and the 50X parfocality will be pursued with the 
vendor prior to making repairs. I am now in touch with
Carol Horn @ Cambridge re: their terrible service.


>>> 'reichert' problem from bob -- Mon Aug 14 07:46:57 1989 <<<

Lou Fishbine of Cambridge Inst. called to get our address so he can
set up an appointment with field service. They will be looking into
the problem with parfocality of the 50X obj., our filter holder not
being removable and our camera/dirt problem. Users will be notified 
of the appointment date in advance.

>>> 'reichert' problem from bob -- Tue Aug 15 11:56:15 1989 <<<

Problem: Filter holder not removable and 50X-100X not parfocal.

Cause:   Lamp assembly supplied is for Polyvar not Polylight. 
         100X objective not seated correctly.

Solution:The filter holder is removable by unscrewing the lamp
         assembly. It is an excellent light source therefore it
         will not be exchanged and there is little problem per-
         forming the filter removal. The filter holder takes 30mm
	 filters and installing our own will be little
	 problem. 

	 The 100x objective was properly installed and is
	 now parfocal with the 50X within 3-4 microns.

	 The reichert is up ready for use.

>>> 'reichert' problem from bob -- Wed Sep 13 15:13:09 1989 <<<

A modification has been made to the Reichert at the request of
Carl Galewski. The automated rotation is now set up so it is
by-directional except when the objective is in the lowest
magnification, 2X. In the 2X position, only forward motion is
allowed. This ingenious modification is documented in the
master copy of the Reichert manual.

>>> 'reichert' problem from cgoss -- Fri Sep 29 14:49:27 1989 <<<

Problem:  No illumination
Probable cause:  Dead bulb.  No light emmerges from bulb housing.

>>> 'reichert' problem from cgoss -- Fri Sep 29 14:51:26 1989 <<<

No illumination problem reported separately as a problem.

>>> 'reichert' problem from bob -- Fri Sep 29 15:52:53 1989 <<<

The illumination on the reichert works fine. There are also spare available
to users in the reichert drawer. Reviewing cgoss' message reveals the basic
reichert problem to be Charles' lack of syntax. The reichert is up.
>>> 'reichert' problem from galewski -- Thu Oct 26 12:38:15 1989 <<< [No Comment]

>>> 'reichert' problem from galewski -- Thu Oct 26 12:38:34 1989 <<<

Problem:  100x objective does not give centered view when stepping
	  from 50x.  Also image drifts at 100x.  Gives different\
	  view when stepped from 150x

Analysis:  Black tape on objective holder has a seam where drive wheel
	   hits objective holder when viewing through 100x.  Glue ending
	   up on o-ring from both black and white tape.  Improved some
	   on drift problem after increasing tension of drive wheel but
	   difficult to use since image not centered or consistent.

Editorial:  Maybe we can think of something more durable and smoother
	    for the 150x lock-out detector?

				-Carl

>>> 'reichert' problem from galewski -- Thu Oct 26 13:52:06 1989 <<<

At Evan's request I checked my analysis.

Yes releasing the drive wheel and using manual rotation all objectives
centralized and no drifting.

				-Carl

>>> 'reichert' problem from voros -- Mon Oct 30 15:54:10 1989 <<<

Clutch is disengaged, rotation can be done only manually, until
motor rotation is fixed. (It is listed in techjobs.)

>>> 'reichert' problem from evan -- Tue Oct 31 11:34:43 1989 <<<

problem: subtle focus shift
cause: drive wheel hitting seam in black tape of motion detector
solution: increased intensity of IR LED on the optical pickup
         - removed all the black tape
         - put a small piece of reflective tape where needed for sensor
             that does not sit under drive wheel at any lens stop.

reichert lens control is working properly again

>>> 'reichert' problem from jlou -- Sun Dec  3 22:16:07 1989 <<<

The lamp doesn't work. I believe the contact electrodes are too dirty.
                                             Jlou

>>> 'reichert' problem from evan -- Mon Dec  4 11:18:57 1989 <<<

problem: lamp not lighting
cause: pigtail to lamp holder connection burnt
solution: remade connection using clean screw and washer

reichert is up

>>> 'reichert' problem from rolfe -- Thu Dec  7 14:09:22 1989 <<<

Excessive vibration at 500x -- no air in left-front air-cushion.

>>> 'reichert' problem from evan -- Fri Dec  8 16:46:29 1989 <<<

problem: air table puck flat
solution: reset pressure in left-front puck

>>> 'reichert' problem from jiahua -- Tue Feb  6 15:32:43 1990 <<<

The film was clamped in the camera.

>>> 'reichert' problem from harshman -- Mon Mar  5 16:24:08 1990 <<<


the

the front, left cornor of the Reichert air table seems to be down

>>> 'reichert' problem from evan -- Mon Mar  5 16:47:05 1990 <<<

problem: air table out of adjustment
solution: manually triggered lifter valves to get some air into front
pucks.
reichert is up

>>> 'reichert' problem from field -- Thu Mar  8 21:24:18 1990 <<<

Initially, had great difficulty removing film from camera, once
exposed.  As time went on, camera acted as it should.

>>> 'reichert' problem from field -- Wed Mar 14 16:42:19 1990 <<<

Film gets stuck in polaroid attachment 3 out of 3 times.  
Position set to L before trying to remove film every time.

>>> 'reichert' problem from bob -- Thu Mar 15 15:12:42 1990 <<<

The reichert has been repaired and is up. The polaroid back had
a jammed film end in it.
From bob Fri Mar 16 12:06:22 1990
Return-Path: <bob>
Received: by argon.berkeley.edu. (4.1/SMI-4.0)
	id AA24730; Fri, 16 Mar 90 12:06:18 PST
Date: Fri, 16 Mar 90 12:06:18 PST
From: bob (Robert M. Hamilton)
Message-Id: <9003162006.AA24730@argon.berkeley.edu.>
To: voros
Subject: polaroid
Cc: bob, debra
Status: RO


The cost of repair for polaroid mod.545 type 52 backs is $80.00 + parts.
Turn around time is 7 days. 
For reference:
Polaroid Customer Service
3232 West McArthur
Santa Anna, CA 92704
1-800-432-7511

In addition, Polaroid is sending an exploded drawing along
with a aprts list. If we buy parts the # is:
1-800-343-4846

The cost of a new back is $254.00 and available through VW&R Graphics Div.

We currently have 2 spare backs and we can repair the one we
have ourselves once we have the parts list.

All our spares are up at the moment.

Bob
.



>>> 'reichert' problem from richard -- Fri Mar 23 14:51:44 1990 <<<

the motorized objective turret does not work; will not change
magnification

>>> 'reichert' problem from evan -- Fri Mar 23 15:52:02 1990 <<<

problem: motor drive for objective turntable not working
cause: someone loosened the turntable assembly lock down knob
solution: returned turntable assembly to set position are retightened 
lockdown knob.  This knob is located on the left side of the throat
just above and to the back of the turntable. This knob should not be
touched by users.  It is a critical setting for parfocal alignment of
the objectives.

reichert is up

>>> 'reichert' problem from hebert -- Mon Apr  9 16:00:53 1990 <<<

The stage lock isn't working.

>>> 'reichert' fix from evan -- Tue Apr 10 14:07:39 1990 <<<

problem: stage lock not working
solution: adjusted stage lock assembly.
reichert is up

>>> 'reichert' problem from moroney -- Fri May 11 19:47:08 1990 <<<

Objective changer occasionally slips and does not change objective
(happened three times or so during this session)

Protection from changing between 2x and 150x objective fails most of the
time, only worked once, I tried it about 10 times.  Maybe the tape is
dirty?  I'm not sure.

No other problems.

>>> 'reichert' fix from evan -- Mon May 14 15:46:06 1990 <<<

problem: 2x to 150x disable sensor not working
solution: repositioned sensor so it 'reads' metalic tape

reichert is up